The SerialEM Script Repository

Check grid

Edit Delete

Details

Name Check grid
Permalink https://serialemscripts.nexperion.net/script/28
Category Other Data Acquisition
Language SerialEM Scripting
Description Quick survey of TEM grid.
Author Jason de la Cruz

Version History

Version 1.0

Download

Upload date June 30, 2017
Download count 642
SerialEM version ≥ 3.7

            
	    	
	    		
	    

Documentation

Modify settings in section below. Set "setIntensityCmd" to either "SetIlluminatedArea" for FEI Titan series or "SetPercentC2" for all other microscopes. Decide on condenser beam illumination (aperture, spot size); enter desired spot size in settings section. Then set "lowMag" to lowest mag with unobstructed view of TEM grid (i.e., the mag you use for whole-grid atlas), adjust intensity for optimum signal, center beam shift, then record intensity value in "lowMagIntensity".

When ready to start, open file to save autodoc metadata (with .mdoc or .idoc) then run the script. Close the file when done.

Please check that beam is centered in low mag before continuing.
Ensure all checkboxes in "Script > Controls..." dialog are unchecked to avoid "low counts" errors.